| PTFE Filter Cartridges | |
![]() |
0.2 Micron Absolute Filter Cartridges |
|
|||||||||||||||||||||||||
| The PTFE Pharmaceutical Liquid
Bacterial Retentive (LBR) cartridge is validated to produce sterility through
retention of Brevundimonas diminuta per HIMA guidelines. PTFE exhibits high air
and liquid flow rates as well as being a sterilizing grade filter that uses
expanded PTFE membrane. Users expecting the ultimate flow rate performance for
liquid or air sterilization will find the Proflow Il LER to be their preferred
choice. PTFE cartridges are manufactured utilizing a certified Class 10,000 cleanroom that operates under guidelines of the FDA's Good Manu-facturing Practices for Medical Devices All modules are flushed with high purity, non-pyrogenic deionized water, 100% integrity tested and serialized to provide complete product traceability. A Certificate of Quality accompanies each completed cartridge detailing all lot, audit and general testing requirements.
|
|||||||||||||||||||||||||
| TYPICAL FLOW RATES AIR: 44 scfm/psid (18 slpm/mbar) LIQUID: 1.9 gpm/psid (1O4 Ipm/bar) MATERIALS OF CONSTRUCTION
STEAM STERILIZABLE AND SANITIZATBLE Cartridges may be Steamed or Autoclaved for over 100 one-hour cycles @ 275ºF (145ºC). Cartridges may also be chemically sanitized in place using common sanitizing agents. Please contact SWT for detailed procedures. BACTERIA RETENTION Bacteria retention greater than 107/cm² per HIMA methodology for 0.2 micron sterilizing grade filters using Brevundimonas dominuta. |
EFFECTIVE FILTRATION AREA 9.3 ft² (0.86 m²) MAXIMUM OPERATING DIFFERENTIAL PRESSURE Forward: 8O psid (5.5 bar) @ 75ºF (24ºC) TRACEABILITY Each PROFLOW II LBR cartridge is given a unique serial number to provide immediate product traceability. INTEGRITY TEST VALUES (Tested in 60% Isopropanol / 40% Deionized Water)
|
||||||||||||||||||||||||